GC Image Users' Guide
Back to FAQ index
Questions about High Resolution MS in GC Image.
Answers to questions about High Resolution MS in GC Image.
- How do I add a new calibrant to Calibrant Database?
- Choose Configure Menu > Mass Calibration > Mass Calibration Settings...
- Select Calibrant Database tab
- Press New, to open Calibrant Editor
- Fill in name and calibration ions data
- Press OK, to add the new calibrant to database
- How do I edit a calibrant in Calibrant Database?
- Choose Configure Menu > Mass Calibration > Mass Calibration Settings...
- Select Calibrant Database tab
- Select a calibrant
- Press Edit, to open Calibrant Editor
- Update calibrant data
- Press OK, to save the change
- How do I delete a calibrant from Calibrant Database?
- Choose Configure Menu > Mass Calibration > Mass Calibration Settings...
- Select Calibrant Database tab
- Select a calibrant
- Press Delete, to remove calibrant from the database
- How do I export a calibrant from Calibrant Database?
- Choose Configure Menu > Mass Calibration > Mass Calibration Settings...
- Select Calibrant Database tab
- Select a calibrant
- Press Export
- Choose file name and location
- Press OK, to export the calibrant
- How do I import a calibrant into Calibrant Database?
- Choose Configure Menu > Mass Calibration > Mass Calibration Settings...
- Select Calibrant Database tab
- Press Import
- Choose an exported calibrant file
- Press OK and a preview dialog is open
- Press OK and then enter a name for the calibrant
- Press OK, to import calibrant to the database
- How do I create an internal calibrator?
- Open MS Viewer for a known reference spectrum
- Press Mass Calibration in toolbar, to open calibrator editor
- Press Load to choose a calibrant from database

- Assign ToFs
- Press Auto Assign ToFs to automatically centroid and assign matched calibration
ions
- Use Manually Assign ToFs to interactively assign selected calibration
ions
- Choose Calibration Function Type and Polynomial Order if applicable
- Press Compute, then the new calibration function is shown below the current calibration function.
- Examine calculate mass accuracy and peak shape to remove ill-formed peaks and repeat Step 4-6.
- Press OK to add calibrator to the calibration model. This calibrator will
now be part of the model and used to calibrate spectra
- How do I assign ToFs to calibrator ions automatically?
- Open Calibrator Editor
- Select rows in table to be updated. (NOTE: If no rows selected, all rows will be
updated)
- Press Auto Assign ToFs to automatically assign ToFs based on Mass Calibration settings.
(NOTE: Any ion that could not be found in the spectrum will not be changed.)
- How do I assign ToFs to calibrator ions manually?
- Open Calibrator Editor
- Select a single row in calibrator table
- Select a single peak in the MS Viewer
- Press Manually Assign ToFs to centroid the peak based on Mass Calibration settings and
assign the calculated centroid to the selected row
- How do I view a calibrator in the calibration model?
- Choose Configure Menu > Mass Calibration > Current Calibration Model
- Select Analysis Calibration tab
- Select desired calibrator from the table
- Press View in toolbar, to show a quick view of the the calibrator's table and
graph
- How do I edit a calibrator in the calibration model?
- Choose Configure Menu > Mass Calibration > Current Calibration Model
- Select Analysis Calibration tab
- Select desired calibrator from the model
- Press Edit in toolbar, to launch MS Viewer associated with this calibrator
- How do I delete a calibrator from the calibration model?
- Choose Configure Menu > Mass Calibration > Current Calibration Model
- Select Analysis Calibration tab
- Select desired calibrator from the model
- Press Delete in toolbar, to remove the calibrator from the model
- How do I export a calibrator table?
- Choose Configure Menu > Mass Calibration > Current Calibration Model
- Select Analysis Calibration tab
- Press Export button, to save the user-defined calibrator table to a file
- How do I import a calibrator table?
- Choose Configure Menu > Mass Calibration > Current Calibration Model
- Select Analysis Calibration tab
- Press Import button, to import a previously exported calibrator table (NOTE:
the new calibrators will be appended to the model.)
Back to FAQ index